Product Highlights
The RX-30, RX-31, and RX-33 represent the pinnacle of energy measurement accuracy, precision, and stability. Engineered for demanding metrology applications, the RX-33 provides industry-leading 0.01% accuracy, long-term measurement stability, and full NIST-traceability, making it ideal for meter test systems, calibration labs, and standards laboratories.
With support for both single-phase and polyphase configurations, advanced measurement capabilities, and seamless integration with WECO test equipment, the RX-30, RX-31, and RX-33 ensure unmatched performance and confidence in every test. Built with RADIAN’s legacy of excellence, they set the standard for reliable and repeatable energy measurement.
Precision Perfected with RADIAN
RADIAN RX-30
The RADIAN RX-30 provides ±0.04% energy accuracy (±0.02% at unity power factor), making it a dependable choice for meter testing, calibration, and precision measurement work. It supports wide voltage, current, power, and frequency ranges while maintaining consistent performance across typical utility and laboratory environments.
Voltage accuracy is ±0.02% and direct AC current accuracy is ±0.028% at TCAL. Power accuracy matches the energy class at ±0.04%. Long-term drift is managed by a stability specification of 0.0016% × √month, and the energy temperature coefficient is 0.0005%/°C, supporting predictable results as conditions change.
Performance Highlights
• Power and energy accuracy: ±0.04% (±0.02% at unity PF)
• Voltage accuracy: ±0.02%
• Current accuracy: ±0.028%
• Stability: 0.0016% × √month
• Energy temperature coefficient: 0.0005%/°C
RADIAN RX-31
The RADIAN RX-31 delivers ±0.02% energy accuracy (±0.01% at unity power factor), a reduced uncertainty class compared to the RX-30. The same accuracy applies to its power measurements, supporting applications that require tighter uncertainty budgets and higher traceability.
Voltage accuracy is ±0.01% and direct AC current accuracy is ±0.014% at TCAL. A stability specification of 0.0004% × √month gives improved control over drift, and the energy temperature coefficient remains 0.0005%/°C.
Performance Highlights
• Power and energy accuracy: ±0.02% (±0.01% at unity PF)
• Voltage accuracy: ±0.01%
• Current accuracy: ±0.014%
• Stability: 0.0004% × √month
• Energy temperature coefficient: 0.0005%/°C
RADIAN RX-33
The RADIAN RX-33 provides ±0.01% energy accuracy (±0.005% at unity power factor), the highest accuracy class in the RX-30 series. It supports advanced calibration laboratories, metrology centers, and manufacturing environments where minimized uncertainty and maximum traceability are essential.
Voltage accuracy is ±0.005% and direct AC current accuracy is ±0.007% at TCAL. Power accuracy matches the energy class at ±0.01%. Long-term drift is controlled by a stability specification of 0.0004% × √month for both power and energy, and the energy temperature coefficient is 0.0005%/°C.
Performance Highlights
• Power and energy accuracy: ±0.01% (±0.005% at unity PF)
• Voltage accuracy: ±0.005%
• Current accuracy: ±0.007%
• Stability: 0.0004% × √month
• Energy temperature coefficient: 0.0005%/°C


Features
- Ultra-High Accuracy – Industry-leading energy measurement precision with long-term stability and reliability.
- Single-Phase & Polyphase Compatibility – Supports meter testing, calibration, and power quality analysis.
- Multiple Accuracy Classes – Available in ±0.04%, ±0.02%, and ±0.01% accuracy levels to match specific testing needs.
- Advanced Measurement Capabilities – Measures voltage, current, power, phase angle, harmonics, and power factor with unmatched consistency.
- Seamless Integration – Works with WECO Meter Test Systems and other calibration equipment for comprehensive testing.
- Durable & Reliable – Designed for long-term operational stability and minimal maintenance.
- Backed by ISO/IEC 17025 accreditation for industry-leading measurement uncertainties and unmatched customer trust for over 35 years.
Exceptional Accuracy, Precision, and Stability
- Communications: RS-232 with RADIAN legacy RD commands, and Ethernet with SCPI commands.
- Programming guide: the application note Controlling RADIAN RX Reference Standards with SCPI, with command scripts and a Python example for automated test systems.
- Programmable pulse outputs for Wh, VARh or VAh, plus a bidirectional pulse/frequency comparator port.
- Reference clock input (10 or 20 MHz) and an optical sensor port.
- Storage: 2 GB internal memory, plus SD card (up to 32 GB) and USB.
- Display: a built-in 4.3-inch color touch screen, with an optional external touch screen display.
Which RX Standard Is Right for You?
All three RX models share the same three-phase platform, functions and connections. They differ in accuracy class and operating environment. For a single-phase primary transfer standard in the lab, see the RADIAN RD-22.
| Model | Energy accuracy | Operating temperature | Best for |
|---|---|---|---|
| RADIAN RX-30 | ±0.04% | -20 to 65 °C | Meter testing and calibration in the shop or field |
| RADIAN RX-31 | ±0.02% | -20 to 65 °C | Meter shops and calibration work that need tighter uncertainty |
| RADIAN RX-33 | ±0.01% | 0 to 40 °C | Calibration labs, metrology centers and manufacturing |
Energy accuracy is at 0.5 power factor.
Technical Specifications
Accuracy
| RX-30 | RX-31 | RX-33 | |
|---|---|---|---|
| Power and energy, 0.5 lag to 1 to 0.5 lead | ±0.04% | ±0.02% | ±0.01% |
| Power and energy at unity power factor | ±0.02% | ±0.01% | ±0.005% |
| Power and energy, 5 mA to <20 mA | ±0.08% | ±0.04% | ±0.02% |
| Voltage, 30 V to 600 V (525 V at 50 Hz) | ±0.02% | ±0.01% | ±0.005% |
| Current, 20 mA to 200 A | ±0.028% | ±0.014% | ±0.007% |
| Current, 5 mA to <20 mA | ±0.056% | ±0.028% | ±0.014% |
| Phase angle | 0.008° | 0.005° | 0.003° |
| Repeatability | 0.0001% ÷ √(test time in seconds) | 0.0001% ÷ √(test time in seconds) | 0.0001% ÷ √(test time in seconds) |
Accuracy is stated in percent of reading, not percent of scale. Repeatability is already included in the accuracy figures. Power and energy accuracy applies to tests of 1 second or longer.
Stability and Temperature
| RX-30 | RX-31 | RX-33 | |
|---|---|---|---|
| Stability, power and energy (× √month) | 0.0016% | 0.0008% | 0.0004% |
| Stability, voltage (× √month) | 0.0008% | 0.0004% | 0.0002% |
| Stability, current (× √month) | 0.0012% | 0.0006% | 0.0003% |
| Temperature drift, power and energy | 0.0005%/°C | 0.0005%/°C | 0.0005%/°C |
| Temperature drift, voltage | 0.00015%/°C | 0.00015%/°C | 0.00015%/°C |
| Temperature drift, current | 0.0003%/°C | 0.0003%/°C | 0.0003%/°C |
| Specified temperature range | −20 to +65 °C | −20 to +65 °C | 0 to +40 °C |
| Operating temperature range | −20 to +65 °C | −20 to +65 °C | −20 to +65 °C |
| Storage temperature | −40 to +80 °C | −40 to +80 °C | −40 to +80 °C |
| Relative humidity | 95% non-condensing | 95% non-condensing | 95% non-condensing |
Harmonics and Measurement
| RX-30, RX-31, and RX-33 | |
|---|---|
| Measurement bandwidth | 3,840 Hz (to the 64th harmonic at 60 Hz) |
| Fundamental frequency | 45 Hz to 75 Hz |
| Sampling | 20-bit, 1 million samples per second |
| Measurement modes | Broadband (all harmonics) or fundamental only |
| VAR methods | All ANSI and IEC VAR algorithms |
| Metrics | Watts, Wh, VA, VAh, VAR, VARh, totalized values, volts, amps, phase angle, power factor, and frequency |
| Voltage input | 30 V to 600 V, 1 MΩ input impedance |
| Current input | 1 mA to 200 A on one range, so there is no range switching across the full span; input impedance less than 500 µΩ |
Automation and Interfaces
| RX-30, RX-31, and RX-33 | |
|---|---|
| Integration time | 0.5 seconds by default, adjustable in 0.5-second steps in PCSuite. Integration time is how long each reading is measured; it is separate from the test time. |
| Ethernet | 1 port, SCPI commands (see the SCPI application note) |
| RS-232 | 1 port, RADIAN legacy RD commands |
| USB | 2 ports |
| Pulse outputs | 4 programmable bidirectional ports, up to 4 MHz |
| Reference clock input | 10 MHz or 20 MHz |
| Warm-up | Less than 60 seconds |
Physical, Safety, and Calibration
| RX-30, RX-31, and RX-33 | |
|---|---|
| Power supply | 100 V to 240 V, 50 Hz or 60 Hz, about 35 VA |
| Dimensions (D × W × H) | 190 × 394 × 114 mm |
| Weight | 6.8 kg (15 lb) |
| Maximum altitude | 2,000 m |
| Enclosure | IP30 |
| Conformity | CE; protection class I (EN 61140) |
| Overvoltage category | CAT IV 600 V, voltage and current measurement |
| Recommended calibration interval | 1 year |
| Calibration | NIST-traceable; ISO/IEC 17025 accredited calibration with stated uncertainties available |
| ISO/IEC 17025 calibration uncertainty | Voltage ±0.003%; current ±0.005%; power and energy ±0.005%; frequency ±0.000005% |
| Documents | Sample calibration reports: RX-30: ISO/IEC 17025 · ANSI/NCSL Z540-1 RX-31: ISO/IEC 17025 · ANSI/NCSL Z540-1 RX-33: ISO/IEC 17025 · ANSI/NCSL Z540-1 ISO/IEC 17025 scope of accreditation |
| Brochures | RX-30/RX-31 Brochure and RX-33 Brochure |
RX Reference Standards FAQ
What is the RX operating range?
What does an RX standard measure?
How does an RX standard connect to a test system?
Can I use an RX standard in the field?
How often should an RX standard be recalibrated?
Related Resources
Strength in Numbers: Reducing Uncertainty with Three RD-22 Primary Transfer Standards


Can't Find What You're Looking For?